12 World Precision Instruments
usage, the batteries will operate satisfactorily for two months or more. Replace
batteries when required or annually. Model A365R requires 16 Nickel-Metal Hydride
9-volt batteries (7.2volt); type IEC-6F22 or equivalent is recommended. Alkaline
batteries in model A365D cannot be replaced with NiMH rechargeable batteries as
model A365D is not equipped for charging. Do not use the higher voltage 9.6.V
rechargeable cells in this instrument.
Procedure
Turn the POWER switch to OFF. Remove four screws on the bottom of the instrument
case. Gently remove the entire internal instrument assembly from the outer case
by sliding the front panel forward. Unscrew and the remove padded cover plate
over the battery assembly. Remove this plate by lifting up and moving out the right
side of the instrument. When removing the batteries work from the front row fi rst
on either side and fi rmly pull up, but without tilting the batteries excessively as not
to damage the battery clips. Insert the new batteries with fi rm but moderate force
to seat the battery in the clips. Match with care the male and female connectors
on the battery tops to the mating connectors on the battery printed circuit board.
When all batteries have been installed, reattach the battery cover plate and test the
instrument by repeating the “Quick Instrument Tests” outlined above. Nickel-Metal
Hydride rechargeable batteries may require an initial 16-hour charge, depending
on their condition when purchased. Reinsert the instrument assembly into the
outer case and secure the bottom screws. Take care to push the copper contacts
down when inserting the case. Nickel-Metal Hydride rechargeable batteries will
discharge by themselves on the shelf. If the A365 is to be stored for a length of
time (up to 3 months), then test the unit and either recharge the batteries or replace
them as necessary. For long term storage remove the batteries from the A365.
Nickel-Metal Hydride rechargeable batteries should last approximately one to two
years with proper care. The replacement batteries should be chosen from the same
manufacturer (and the same manufacturing lot number, if possible) to ensure equal
levels of drain.
Expected Battery Life
9V Alkaline main high voltage stack: 57 hours at 100% duty cycle 10 mA load;
428h @ 1 mA, 1200h @ 100μA.
7.2 V NiMH main high voltage stack: 14 hours at 100% duty cycle 10 mA load;
100h @ 1mA; 306h @ 100μA.
Typical duty cycles are 1 to 5% (At a full load of 10mA and at a 1% duty cycle the
A365 can last approximately 1000hours on alkaline cells, on NiMH this duration
is approxi-mately 260 hours, both are due to the 9V power circuit discharging fi rst.
Control circuit (powered by one 9V or 7.2V battery):
Power on and not stimulating: 1000 hours ( 9 V alkaline), 260 hours (7.2 V NiMH)